Intuitive Surgical Battery Box of the da Vinci STM and da Vinci SiTM Surgical Systems 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Device Technologies Australia Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2014-RN-00892-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2014-09-02
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    Patient side cart battery boxes returned to intuitive surgical from the field have revealed, in some rare charging conditions, the battery has been found to heat sufficiently to cause the battery case to bulge. a battery case that has bulged has the potential to release some of its internal gases to the surrounding air. the da vinci stm and da vinci sitm surgical systems are designed to detect overheating batteries and issue a system warning, error code 808 is displayed on the surgeon console and patient side cart monitors. when this event is detected, the system disables the charging of the battery until it cools down.
  • 조치
    Users are advised that they may continue to use their da Vinci S, Si and Si-e Patient Side Carts. However, if users receive an Error Code 808, or experience a sulfur odour while using da Vinci STM or da Vinci SiTM Surgical System, they are advised to contact their Device Technologies representative, following standard troubleshooting process. For product already in use, Device Technologies is utilising standard repair and preventative maintenance programs to monitor battery performance and perform battery replacement as needed. This action has been closed-out on 06/06/2016.

Device

  • 모델명 / 제조번호(시리얼번호)
    Intuitive Surgical Battery Box of the da Vinci STM and da Vinci SiTM Surgical SystemsProduct number for the battery box: 371268-02 Manufactured between 22 November 2005 and 09 July 2012ARTG Number: 97348
  • Manufacturer

Manufacturer