Invacare Birdie Lifters (Patient lifters)Lifter Birdie (180KG-EL-JC) 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Invacare Australia Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2016-RN-01287-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2016-10-06
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    Invacare are updating the instructions for use with respect to correct assembly of the d-clip that connects the actuator to the birdie lifter’s arm. as a result of some reports overseas where the actuator’s d-clip becomes damaged and the lifter arm falls down, invacare have become aware that if the d-clip is always placed backwards and pointing down towards the actuator (incorrect assembly), it can start to become damaged after a number of cycles. if the d-clip becomes damaged to the extent that the lifter arm falls down then there is potential for a serious patient injury to occur. when the clip is in the wrong position the bending (damage) of d-clip only happens if it is touching the actuator backwards and at its maximum height. the clip will also bend severely when over 140kg is used with the device. the complete clevis pin (quick release pin) doesn’t detach immediately from the actuator when the d-clip becomes damaged - the pin stays in the arm’s plates that hold the actuator.
  • 조치
    Invacare is advising customers that the following should be checked on each Birdie Lifter. 1. The actual state of the D-clip - does it appear bent or broken? If the D-clip is damaged or broken, Invacare should be contacted and they will provide replacement stock. 2. The actual position of the D-clip in the lifter. If it is in the correct position, there is no need to change it. If it is assembled in the wrong position, the D-clip should be removed and reassembled according to the directions that are being provided. Revised instructions for use are avaliable.

Device

  • 모델명 / 제조번호(시리얼번호)
    Invacare Birdie Lifters (Patient lifters)Lifter Birdie (180KG-EL-JC)Item Number: 3000503.E3B00-0125Lifter Birdie (170KG EL)Item Number: 3000502.E3B00-0125Lifter Birdie CMPT (150-ML-JC)Item Number: 3000513.D3B02-0125Lifter Birdie Compact (150KG ML)Item Number: 3000512.D3B02-0125Lifter Birdie MAN (170KG-ML-JC)Item Number: 3000503D3B02-0125ARTG Number: 166395
  • Manufacturer

Manufacturer