Invivo DynaCAD Breast and Prostate with software versions 3.4 and 3.5 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Philips Electronics Australia Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2018-RN-00163-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2018-02-21
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    A problem was uncovered which causes images processed with pharmacokinetic (pk) results and/or dynacad calculated apparent diffusion coefficient (adc) maps have incorrect colour overlay results as well as in the case of the adc images, incorrect values reported. the defect exists on any dynacad client system running software versions 3.4 or 3.5.
  • 조치
    Philips is advising users to discontinue use of the Ktrans, Kep, and iAUGC colormaps when assessing studies from a remotely connected DynaCAD client computer. It is further advised that users refrain from using the colormaps and values derived from ADC maps calculated by DynaCAD. ADC maps originating natively from the MRI system should be used as an alternative as these values are unaffected. Philips will provide a v4.0 software update for the affected software versions as a permanent correction.

Device

  • 모델명 / 제조번호(시리얼번호)
    Invivo DynaCAD Breast and Prostate with software versions 3.4 and 3.5Multiple serial numbers affectedARTG Number: 230620
  • Manufacturer

Manufacturer

  • 제조사 모회사 (2017)
  • Source
    DHTGA