iPlan RT Dose / iPlan RT Versions 4.5.3 and 4.5.4, when used in combination with the optional function Dose Planning for Agility Multileaf-Collimator (MLC) 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Brainlab Australia Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2016-RN-01317-1
  • 사례 위험등급
    Class I
  • 사례 시작날짜
    2016-10-07
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    The agility mlc controller automatically adds a physical gap of 1mm (about 3mm projected to the isocenter) between opposing closed leaf pairs. therefore, leakage radiation through this gap must be correctly considered during treatment planning and dose calculation. due to a software inconsistency, the brainlab pencil beam dose calculation algorithm does not correctly regard and represent this gap if all of the following conditions are met:1. iplan rt dose v. 4.5.3 or 4.5.4 is used.2. the planning option for agility mlc is used for a treatment plan.3. the machine profile (dose profile entered in the brainlab treatment planning software) for the used agility mlc and linear accelerator combination contains a leaf shift static value different than 0mm.If the above described conditions are met, only specific plans irradiating multiple separate target volumes simultaneously with a single isocenter are affected.
  • 조치
    Users are advised to immediately check the Leaf Shift Static value in all the machine profiles in iPlan RT Dose for an Agility MLC: Ensure that this value is set to “0 mm”. If another value is found, please change this specific value to 0 in order to permanently resolve the described potential issue. The users are also reminded of the following information which is in the current IFU: - For any new or changed machine profiles, it is highly recommended to perform a dose verification (using a phantom) before clinical use, to ensure adequate implementation and accuracy of dose calculation. - Patient-specific Quality Assurance of a treatment plan and verifying the dose distribution (in a suitable phantom) is highly recommended before irradiating the patient. Brainlab will be offering a software solution and will contact affected customers starting November 2017 to implement the update free of charge.

Device

  • 모델명 / 제조번호(시리얼번호)
    iPlan RT Dose / iPlan RT Versions 4.5.3 and 4.5.4, when used in combination with the optional function Dose Planning for Agility Multileaf-Collimator (MLC)ARTG Number: 121114
  • Manufacturer

Manufacturer