James Leckey Design Mygo Stander 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Otto Bock Australia Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2016-RN-00854-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2016-06-29
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    It has been identified that the initial design of the angle adjustment mechanism for the stander allowed the user to lock out the safety pull pin on the frame and adjust the angle of the stander using one hand. if the operator’s hand slipped off the turning handle when a patient was in the frame then the handle could “self-wind” and spin rapidly as the stander returns to the upright position. whilst there is no risk to the patient in the frame, the fast spinning handle does pose a risk of injury to the operator if they try to take hold of the handle.These issues only impact devices manufactured between january 2012 and august 2013 as the design of the pull pin safety mechanism was changed in august 2013. it is no longer possible to lock the pull pin out and two hands are now required to adjust the stander angle.
  • 조치
    Otto Bock is arranging for distributors to contact users of the affected devices and arrange for replacement of the pull-pin on the device with one that cannot be locked by the operator. The replacement safety pull pin will require the user to hold the pull pin with one hand while turning the adjustment handle with the other hand. If their hand slips off the turning handle, they can simply release the pull pin and the handle will stop winding. For more information, please see https://www.tga.gov.au/alert/james-leckey-design-mygo-stander .

Device

  • 모델명 / 제조번호(시리얼번호)
    James Leckey Design Mygo StanderMygo Stander Size 1Product Code: 143-1600Mygo Stander Size 2Product Code: 143-2600 Serial Number range: LEC/087023 to LEC/108416Manufacture dates: January 2012 to August 2013ARTG Number: 130629
  • Manufacturer

Manufacturer