Laerdal Compact Suction Unit 4 (LCSU 4) (portable suction unit intended for field and transport use) 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Laerdal Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2016-RN-01529-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2016-11-24
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    Laerdal medical has become aware of a problem related to the on/off functionality of the lcsu 4. the on/off functionality is controlled by electronic components and one of these components might be damaged. the damage occurs over time when the unit is connected to a power source (battery or external power source), but is not related to frequency of use. a damaged component will result in failure of the on/off functionality whereby the lcsu 4 cannot be turned off, and it may turn on by itself. the unit can be turned off by removing the battery or the external power source.
  • 조치
    Laedral will be providing replacement units. In the interim, Laedral is advising users to perform a test to identify if their unit has failing On/Off functionality, by turning the unit on then turning it off again. If the unit cannot be turned off using the On/Off button, the unit contains a damaged component and the device should be taken out of service. If the unit passes the test, the device can continue to be used until the replacement unit is received, but the following extra safety measures should be implemented: - Store the LCSU 4 unit without the battery connected. - Perform routine checks of the On/Off functionality by turning the unit on and off using the On/Off button. If the On/Off functionality fails during clinical use, the unit can continue to be used so long as it has power. The unit can be turned off by removing the battery or the external power source.

Device

  • 모델명 / 제조번호(시리얼번호)
    Laerdal Compact Suction Unit 4 (LCSU 4) (portable suction unit intended for field and transport use)LCSU 4, 300ml Complete unitItem Number: 88006140LCSU 4, 800ml Complete unitItem Number: 88005140 Units manufactured from 1 May 2015 to 28 October 2016 that do not have 'RW' on labelARTG Number: 205931
  • Manufacturer

Manufacturer