Leica M525 F40 Surgical Microscopes 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Leica Microsystems Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2017-RN-00771-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2017-08-29
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    Leica microsystems has identified that the instructions for use (ifu) and warnings are not always followed by users of the leica m525 f40 surgical microscopes, in particular for transportation and positioning in the operating room (or). due to this, additional waning stickers have been introduced for the leica m525 f40, in addition to a quick reference guide for transporting and position in the or which can be attached to the instrument.Using the m525 f40 improperly during transport and positioning can have an adverse effect on the stability of the device. in extreme cases, improper use during transport and positioning can cause tilting of the device. leica microsystems has received two reports where a m525 f40 started tilting due to improper handling.
  • 조치
    1. Leica Microsystems is sending warning stickers and “quick reference guides for transport and positioning in the OR”. 2. Users are to arrange the fixation of the warning stickers on the instruments, inform or personnel accordingly and fill out the supplied acknowledge form.

Device

  • 모델명 / 제조번호(시리얼번호)
    Leica M525 F40 Surgical MicroscopesMaterial Number: 10448096Serial Numbers: 010708001, 010906002, 060505002, 080508001, 100105001, 100608002, 110608001, 180205002, 180410001, 230507001, 231209001, 240609001, 260609001, 290508001, 301106002ARTG Number: 137683