Leksell Gammaknife Perfexion and Leksell Gammaknife Icon (used for stereotactic irradiation of head structures) 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Elekta Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2016-RN-00410-1
  • 사례 위험등급
    Class I
  • 사례 시작날짜
    2016-04-06
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    There is a potential for incorrect mounting of the frame adapter to the coordinate frame. the latches of the frame adapter can be locked even if the locating pins of the frame adapter are not inserted into the corresponding holes in the coordinate frame. if the frame adapter is mounted incorrectly to the coordinate frame and this is not noticed by the user before treatment is started, dose will be delivered in the wrong place by up to 2cm.
  • 조치
    Elekta is advising users that the frame adapter is safe to use provided that the Instructions for use are followed and conducted by a trained professional. It is of the highest importance that the frame adapter is mounted to the coordinate frame as described in the applicable Instruction for Use. The locating pins of the frame adapter shall be inserted into the corresponding holes in the coordinate frame before the three latches are locked. Also make sure that only trained personnel are involved in mounting the frame adapter. A field corrective action will be implemented to facilitate a correct mounting of the frame adapter.