Lumenis FemTouch Delivery System (used with AcuPulse laser system and AcuScan120 microscanner for gynecological treatments) 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Lumenis Australia Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2016-RN-01331-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2016-11-07
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    Lumenis have learned that the cleaning instructions provided in femtouch operator’s manual (um-10065340 rev. b) might be ineffective for a fully assembled device and may lead to a potential for cross-contamination. lumenis initiated a re-verification of the cleaning protocol, and the results showed the current protocol as stated in the instructions for use (ifu) were not effective for an assembled handpiece (tip). this issue affects all femtouch hand pieces and kits. lumenis has not received any reports of injury or harm associated with this potential risk.
  • 조치
    Lumenis is asking customers to disassemble the FemTouch tip prior to its cleaning process. This will remove cleaning challenges and allow the device to be effectively cleaned and sterilised using the protocol in the labelling, thereby reducing the potential for cross-contamination.

Device

  • 모델명 / 제조번호(시리얼번호)
    Lumenis FemTouch Delivery System (used with AcuPulse laser system and AcuScan120 microscanner for gynecological treatments)FemTouch KitModel Number: AC-1007415FemTouch Tip (supplied in FemTouch Kit)Part Number: SPSA-20002590ARTG Number: 204126
  • Manufacturer

Manufacturer