LX3 Microscope Floor Stand used with the LuxOR Ophthalmic Microscope and LuxOR Ophthalmic Microscope with Q-VUE 3D Assistant Visualisation (used during ophthalmic surgical procedure for cataract, retina and cornea) 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Alcon Laboratories Australia Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2015-RN-00757-1
  • 사례 위험등급
    Class I
  • 사례 시작날짜
    2015-08-21
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    Alcon is undertaking a recall for product correction of all lx3 microscope floor stands after receiving reports concerning loss of functionality due to issues with the cabling between the power, host and illumination modules, potentially causing the system to display a system message (4xx), and/or resulting in illumination failure. alcon will retrofit the cable design on any impacted lx3 microscope floor stands. loss of system functionality during surgery due to issues within the microscope modules or illumination failure will result in the loss of visualisation by the surgeon during the procedure, which can lead to the potential for patient injury. this could include scleral/corneal trauma, intraocular tissue damage, capsule injury, retinal tear/detachment, endothelial cell damage, corneal oedema or intraocular bleeding.
  • 조치
    Alcon is retrofitting all affected devices with a new cable design. Alcon is recommending that if users choose to continue to use the affected devices before the correction has been performed, to ensure an alternative microscope and light source is available. This action has been closed-out on 17/08/2016.