Mako Onlay Insert Extractor (part of the Restoris MCK (Multi Compartmental Knee) Unicondylar System) 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Stryker Australia Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2018-RN-00280-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2018-04-03
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    Stryker has received a report that a hinge pin disassociated from the mako onlay insert extractor. in that reported case, the disassociated hinge pin was discovered prior to the device being used in surgery and a backup device was available and subsequently used to complete the procedure. in the event of an undetected disassociated hinge pin, the potential hazards include the hinge pin being left in the wound or excessive metal wear giving rise to debris. to date, there have been no injuries reported as a result of this issue.
  • 조치
    Stryker is advising users to inspect stock and return any remaining units of the affected lot numbers.

Device

  • 모델명 / 제조번호(시리얼번호)
    Mako Onlay Insert Extractor (part of the Restoris MCK (Multi Compartmental Knee) Unicondylar System)Item Number: 160430Lot Numbers: 19020414, 19090616, 19090915, 19100616, 19100915, 19110317, 19110616, 19110915, 19130315, 19140315, 19451016,19471016, 19490515, 19500515, 19510515, 19520515, 19530515, 26051212, 26070512, 26080913, 26130512, 26170513, 26201111, 26290412, 26440912ARTG Number: 223107
  • Manufacturer

Manufacturer

  • 제조사 모회사 (2017)
  • Source
    DHTGA