Mandible Distractor and Proximal Footplate (used to stabilise bone and lengthen for the correction of congenital deficiencies or post-traumatic defects) 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Johnson & Johnson Medical Pty Ltd t/a DePuy Synthes 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2016-RN-00806-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2016-06-15
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    The fastener on the mandible distractor may become prematurely separated from the proximal foot plate. in the event that the mandible distractor becomes disengaged from the proximal foot plate during device implantation, surgical delay may occur if another device is not available in the surgical suite. if the mandible distractor becomes disengaged from the proximal foot plate after procedure, a repeat surgery may be required to replace the mandible distractor/foot plates. there may be a risk of local infection as a result of repeat surgery caused by the disengagement since the product is in the oral cavity.
  • 조치
    Johnson & Johnson Medical is advising users to inspect stock and quarantine any remaining product for return. This action has been closed-out on 30/01/2017.

Device

  • 모델명 / 제조번호(시리얼번호)
    Mandible Distractor and Proximal Footplate (used to stabilise bone and lengthen for the correction of congenital deficiencies or post-traumatic defects)Part Numbers: 487.962, 487.963, 487.964, 487.965, 487.974, 487.975All lots affectedARTG Number: 183673
  • Manufacturer

Manufacturer