MAQUET operating table accessories with Accessory Adaptor 600525A0 Manufactured from 2002 onwards 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Maquet Australia Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2016-RN-01225-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2016-09-15
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    The accessory adaptor (600525a0) is designed for the mounting of maquet accessories with the dovetail guide interface immediately before, during and after surgical interventions as well as for examination and treatment. the accessory adapter is used with the magnus table back plates to attach the horseshoe head rest or the skull clamp. maquet has been notified of a case where a head rest was levered out of the dovetail profile of the accessory adaptor. it was identified that the raw material of the accessory adaptor did not meet maquet specifications. while the accessory adapter does carry the specified weight of 11kg at a distance of 300mm, the tensile and flexural strength of the material was less than specification.There is a very rare possibility should the attached accessory release from the adapter without warning during a surgical procedure that the patient may sustain a cervical spine injury. to date no incident has been reported in which a person was injured.
  • 조치
    MAQUET will be contacting customers to provide a new replacement Accessory Adaptor (600525A0). Until the replacement has been provided, MAQUET is advising users that Accessory Adaptor (600525A0) should not be used.

Device

  • 모델명 / 제조번호(시리얼번호)
    MAQUET operating table accessories with Accessory Adaptor 600525A0 Manufactured from 2002 onwardsARTG Number: 163023
  • Manufacturer

Manufacturer