MAQUET QUADROX-i Oxygenators (excluding HLS) and Blood Outlet Connectors-QUADROX-i(D) adult, small adult, paediatric and neonatal with Softline and Bioline Coating-PLS-i gas exchanger in PLS sets -PLS gas exchanger in HIT -PLS sets-Quadrox-iR with Softline coating 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Maquet Australia Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2013-RN-00239-1
  • 사례 위험등급
    Class I
  • 사례 시작날짜
    2013-03-15
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    Outlet connector detachment may result in rapid blood loss which could cause serious injury or death. this issue was originally reported under recall rc-2012-rn-00551-3. although the correction implemented for the previous recall reduced the incidence rate of detachment from 0.01% to 0.001%, a patient death potentially attributed to detachment of the outlet connector was reported recently.
  • 조치
    Customers are requested to continue to confirm the integrity of the oxygenator prior to use. The devices should only be used if the connector is secure. Maquet is implementing a stainless steel retainer clamp as a short term correction. A permanent mechanical fitting is currently under development and is expected to be implemented in late 2013. This action has been closed-out on 14/04/2016.

Device

  • 모델명 / 제조번호(시리얼번호)
    MAQUET QUADROX-i Oxygenators (excluding HLS) and Blood Outlet Connectors-QUADROX-i(D) adult, small adult, paediatric and neonatal with Softline and Bioline Coating-PLS-i gas exchanger in PLS sets -PLS gas exchanger in HIT -PLS sets-Quadrox-iR with Softline coating
  • Manufacturer

Manufacturer