MDC PACS R2.3 SP2 HF6 and MDC PACS R2.3 SP2 HF7 (Picture, archiving and communication system) 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Philips Electronics Australia Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2013-RN-01110-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2013-10-29
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    A software error exists that changes the study time sent from modalities when it ends in zero(es) and incorrectly displays it in the pacs work list. the result is that examinations are displayed with a study time of 00:00:00 or 24:00:00 instead of the actual time the study was taken at, resulting in the chronological order of these exams being incorrectly displayed. this incorrect display of study time/ incorrect chronological order may result in patient misdiagnosis.
  • 조치
    Philips is advising users to rely on the study time displayed in DiagNET viewer and not in the PACS client worklist in case studies acquired on the same date. A software update will be provided to permanently correct the issue.

Device

  • 모델명 / 제조번호(시리얼번호)
    MDC PACS R2.3 SP2 HF6 and MDC PACS R2.3 SP2 HF7 (Picture, archiving and communication system)Serial Number: MCD121100431ARTG Number: 207252
  • Manufacturer

Manufacturer

  • 제조사 모회사 (2017)
  • Source
    DHTGA