ME DPHC 0001 deltastream HC, 230V (Cardiopulmonary bypass system heating/cooling unit) 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Lovell Surgical Supplies International Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2016-RN-00362-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2016-04-07
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    There have been reported cases of nontuberculous endocarditis caused by mycobacteria or infections after cardiac surgery, in which hypothermia or hyperthermia devices were used. the manufacturer is updating the cleaning procedure of the device as a precautionary measure. updated ifus are expected to be available by the end of june 2016.
  • 조치
    The manufacturer is advising users to follow the following updated cleaning procedures: · Clean the device every 14 days with Sanosil solution (3% H202). Since there may be an excess pressure during the cleaning process the tank lid must remain open during the cleaning process Furthermore, it is recommended that the water is changed daily and after each patient. · Change least once a year, the hose assembly. · All relevant hypothermia and hyperthermia devices should be operated in a separate room from the operating theatre. If this is not possible, another suitable measure should be found to ensure the risk of infection is effectively minimised. For example, by increasing the disinfection frequency of the surfaces of the devices and changing gloves after cleaning the equipment. · Perform regular checks for contamination of equipment. The manufacturer recommends monthly intervals to determine the quality of the water.

Device

  • 모델명 / 제조번호(시리얼번호)
    ME DPHC 0001 deltastream HC, 230V (Cardiopulmonary bypass system heating/cooling unit)All Serial Numbers affectedARTG Number: 262576