Medi-Trace Cadence Adult Multi-Function Defibrillation Electrodes, Pre-connect (used in conjunction with Automatic External Defibrillators) Medi-Trace Cadence Adult Multi-Function Defibrillation Electrodes 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Covidien Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2015-RN-00803-1
  • 사례 위험등급
    Class I
  • 사례 시작날짜
    2015-08-31
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    This recall is being conducted due to a low-level potential for damage to the wire insulation which was identified during a review of our manufacturing process. no complaints for this issue have been reported from customers. the use of products with this condition may result in a potentially increased risk for either reduced or no patient therapy, arcing of current, sparking, and potential burns to the patient and/or clinician. no patient injuries have been reported regarding this damaged wire insulation issue.
  • 조치
    Users are asked to immediately quarantine and discontinue use of the affected devices and to arrange return as per the instructions supplied. This action has been closed-out on 31/08/2016.

Device

  • 모델명 / 제조번호(시리얼번호)
    Medi-Trace Cadence Adult Multi-Function Defibrillation Electrodes, Pre-connect (used in conjunction with Automatic External Defibrillators)Medi-Trace Cadence Adult Multi-Function Defibrillation ElectrodesItem Numbers: 22550PC, 22770PCLot Numbers: 517521X, 519835X, 516313X, 519124X, 513426X ARTG Number: 186124
  • 의료기기 분류등급
  • Manufacturer

Manufacturer

  • 제조사 모회사 (2017)
  • 제조사 의견
    “If our surveillance systems identify a potential performance issue, our personnel promptly evaluate the problem, including, when appropriate, conducting root cause investigations and internal testing to assess whether the product continues to meet specifications and defined performance criteria,” Medtronic told ICIJ in a statement. “In some cases, based on this evaluation, Medtronic may determine that a recall is necessary.” The company said that it communicates with healthcare providers and/or patients and provide recommendations to address such issues. Medtronic noted that these communications can include letters, emails, calls, press releases, physician notifications and social media postings, as well as informing the FDA and other regulators of the actions.