Medtronic Activa PC, Activa SC, Activa RC, Kinetra and Soletra Implantable Deep Brain Stimulators (DBS) 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Medtronic Australasia Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2015-RN-00673-1
  • 사례 위험등급
    Class I
  • 사례 시작날짜
    2015-07-21
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    Medtronic’s monitoring of reported events, clinical trials and published literature has resulted in labelling updates: warnings:- severe, life-threatening dystonia symptoms, including status dystonicus, during ongoing or loss of therapy may result in respiratory compromise and rhabdomyolysis. rarely, rhabdomyolysis may progress to multi-organ failure & death. - cessation or reduction of stimulation may lead to an increase in seizure frequency or severity of epilepsy. symptoms may return with an intensity greater than was experienced prior to system implant, including the potential for status epilepticus. adverse events: - meningitis, encephalitis or brain abscess resulting from infection - focal oedema to the area around the lead- immediate or delayed intracranial haemorrhage or cerebral infarction which could be symptomatic, or which could result in temporary or permanent neurological injury or death- aseptic intraparenchymal cyst formation around the lead tip.
  • 조치
    The labelling updates further clarify potential risks which have been reported with DBS Therapy. Medtronic is providing the information about the labelling updates to help physicians who are implanting and managing patients to give them visibility into the upcoming changes for labelling and help them with management of their current patients treated with DBS Therapy. Regarding DBS Therapy for Dystonia, physicians are advised to monitor patients receiving DBS Therapy for the above symptoms and emphasise the importance of contacting the physicians if patients experience worsening of seizure frequency or severity. Regarding DBS Therapy for epilepsy, it is important that the patient or caregiver knows how to use the patient programmer in case the neurostimulator is accidentally turned off . This action has been closed-out on 15/08/2016.

Device

  • 모델명 / 제조번호(시리얼번호)
    Medtronic Activa PC, Activa SC, Activa RC, Kinetra and Soletra Implantable Deep Brain Stimulators (DBS)Activa PC Model Number: 37601ARTG Number: 160118Activa SC Model Number: 37603 and 37602ARTG Number: 188034Activa RC Model Number: 37612ARTG Number: 160117Kinetra Model Number: 7428ARTG Number: 134476SoletraModel Number: 7426ARTG Number: 134475
  • Manufacturer

Manufacturer

  • 제조사 모회사 (2017)
  • 제조사 의견
    “If our surveillance systems identify a potential performance issue, our personnel promptly evaluate the problem, including, when appropriate, conducting root cause investigations and internal testing to assess whether the product continues to meet specifications and defined performance criteria,” Medtronic told ICIJ in a statement. “In some cases, based on this evaluation, Medtronic may determine that a recall is necessary.” The company said that it communicates with healthcare providers and/or patients and provide recommendations to address such issues. Medtronic noted that these communications can include letters, emails, calls, press releases, physician notifications and social media postings, as well as informing the FDA and other regulators of the actions.