Medtronic DBS Pocket Adaptor for use with Activa PC (Model 37601) and Activa RC (Model 37612) brain neuro-stimulators 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Medtronic Australasia Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2016-RN-00847-1
  • 사례 위험등급
    Class I
  • 사례 시작날짜
    2016-07-13
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    It has been identified that 16 medtronic neuromodulation dbs system pocket adaptors (0.08% reported rate of occurrence) have been identified with high impedance, resulting in decreased current being conducted through the device. some of the reported cases involved loss of therapy, return of symptoms, rebound effects and potential revision surgery to replace the device. there have been no incidents reported in australia. analysis of the returned devices found the cause of the high impedance to be conductor wire fractures near where the wire exits the neurostimulator connector block. in two cases, the issue was identified during implantation surgery, while the other 14 were identified after implantation and resulted in revision surgery.
  • 조치
    Medtronic is reminding users of the device labelling specific to the handling of pocket adaptors and system integrity checking during implant procedures. This action has been closed-out on 03/03/2017.

Device

  • 모델명 / 제조번호(시리얼번호)
    Medtronic DBS Pocket Adaptor for use with Activa PC (Model 37601) and Activa RC (Model 37612) brain neuro-stimulatorsModel Numbers: 64001 and 64002ARTG Number: 165114
  • 의료기기 분류등급
  • Manufacturer

Manufacturer

  • 제조사 모회사 (2017)
  • 제조사 의견
    “If our surveillance systems identify a potential performance issue, our personnel promptly evaluate the problem, including, when appropriate, conducting root cause investigations and internal testing to assess whether the product continues to meet specifications and defined performance criteria,” Medtronic told ICIJ in a statement. “In some cases, based on this evaluation, Medtronic may determine that a recall is necessary.” The company said that it communicates with healthcare providers and/or patients and provide recommendations to address such issues. Medtronic noted that these communications can include letters, emails, calls, press releases, physician notifications and social media postings, as well as informing the FDA and other regulators of the actions.
  • Source
    DHTGA