Medtronic Reveal LINQ Insertable Cardiac Monitor (ICM) 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Medtronic Australasia Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2016-RN-00644-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2016-06-07
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    In february 2016, medtronic issued a medical device hazard alert communication that indicated medtronic reveal linq insertable cardiac monitors (icms) could experience a performance issue that affects the recommended replacement time (rrt) alert. this original communication named phase 1, also explained that medtronic was awaiting a software update to address this issue.Medtronic has now developed the necessary software and is ready to begin applying this update to all reveal linq icms which constitutes phase 2 of the alert.Once installed, this software update will reset rrt & end of service (eos) status and re-enable wireless transmissions for devices that have experienced premature rrt /eos. the update will also prevent the occurrence of premature rrt alerts due to this issue. refer to the original communication for additional details (rc-2016-rn-00202-1).
  • 조치
    Medtronic will be providing a software update as a permanent correction. For more information, please see https://www.tga.gov.au/alert/medtronic-reveal-linq-insertable-cardiac-monitor. This action has been closed-out on 17/02/2017.

Device

  • 모델명 / 제조번호(시리얼번호)
    Medtronic Reveal LINQ Insertable Cardiac Monitor (ICM) Model LNQ11 ARTG Number: 218791
  • Manufacturer

Manufacturer

  • 제조사 모회사 (2017)
  • 제조사 의견
    “If our surveillance systems identify a potential performance issue, our personnel promptly evaluate the problem, including, when appropriate, conducting root cause investigations and internal testing to assess whether the product continues to meet specifications and defined performance criteria,” Medtronic told ICIJ in a statement. “In some cases, based on this evaluation, Medtronic may determine that a recall is necessary.” The company said that it communicates with healthcare providers and/or patients and provide recommendations to address such issues. Medtronic noted that these communications can include letters, emails, calls, press releases, physician notifications and social media postings, as well as informing the FDA and other regulators of the actions.
  • Source
    DHTGA