Medtronic Spinous Process Clamps 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Medtronic Australasia Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2017-RN-00864-1
  • 사례 위험등급
    Class I
  • 사례 시작날짜
    2017-07-18
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    Medtronic has become aware of instances where spine clamps have been damaged when forced open beyond their intended limits during use and will subsequently not open once attached to a patient’s spinous process. when the spine clamp is forced open beyond its intended limits, a component (captive washer) may break off. the washer could then be inadvertently left behind in a patient’s body if the breakage occurs during the procedure. if the washer is missing from the device, the spine clamp cannot be re-opened after placement on the spinous process.Medtronic has received six reports where unintended removal of spinous process occurred when attempting to detach the spine clamp. the unintended removal of spinous process can lead to damage of adjacent vertebra and cause premature degradation.
  • 조치
    Medtronic is providing an interim workaround for users via the Customer Letter. A new version of the clamps with a design mitigation will be distributed as a permanent correction once available.

Device

  • 모델명 / 제조번호(시리얼번호)
    Medtronic Spinous Process ClampsSpine Referencing InstrumentationModel/Catalogue Numbers9734715 Spinous Process Clamp, Tall9734716 Spinous Process Clamp, Short9734723 Double Spinous Process Clamp, Tall9734724 Double Spinous Process Clamp, Short 9734715K Spinous Process Clamp, Tall9734716K Spinous Process Clamp, Short9734723K Double Spinous Process Clamp, Tall9734724K Double Spinous Process Clamp, Short9735500 Set (contains 9734715 & 9734716)ARTG Number: 125074
  • 의료기기 분류등급
  • Manufacturer

Manufacturer

  • 제조사 모회사 (2017)
  • 제조사 의견
    “If our surveillance systems identify a potential performance issue, our personnel promptly evaluate the problem, including, when appropriate, conducting root cause investigations and internal testing to assess whether the product continues to meet specifications and defined performance criteria,” Medtronic told ICIJ in a statement. “In some cases, based on this evaluation, Medtronic may determine that a recall is necessary.” The company said that it communicates with healthcare providers and/or patients and provide recommendations to address such issues. Medtronic noted that these communications can include letters, emails, calls, press releases, physician notifications and social media postings, as well as informing the FDA and other regulators of the actions.