Medtronic SynchroMed II Pumps 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Medtronic Australasia Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2017-RN-00567-1
  • 사례 위험등급
    Class I
  • 사례 시작날짜
    2017-05-08
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    Medtronic is updating information communicated in july 2011 (rc-2011-rn-00769-3) regarding the failure rate for reduced battery performance in medtronic model 8637 synchromed ii pumps manufactured up to june 2011. updated failure rate information due to this issue:- pumps manufactured mar 2005 through dec 2010: 0.13% cumulative probability for pump failure at 72 months after implant. this rate remains within the failure rate upper bound of 0.2% reported in 2011.- pumps manufactured from jan 2011 through jun 2011: 3.17% cumulative probability for pump failure at 72 months after implant. this failure rate exceeds the upper bound estimate of 0.2% reported in 2011. a patient with a pump exhibiting reduced battery performance may experience return of underlying symptoms and/or withdrawal symptoms. patients receiving intrathecal baclofen therapy are at risk for baclofen withdrawal syndrome, which can lead to a life-threatening condition if not promptly and effectively treated.
  • 조치
    Medtronic is reinforcing the advice provided in 2011. Medtronic does not recommend prophylactic replacement of SynchroMed II pumps with the prior battery design (manufactured before July 2011) because of the estimated low occurrence rates, the presence of pump alarms, and the risks associated with replacement surgery. If Low Battery Reset (critical alarm) or premature Elective Replacement Indicator (non-critical alarm) or End of Service (critical alarm) occurs, replacement surgery should be scheduled as soon as possible. Further ongoing patient management recommendations are detailed in the hazard alert communication provided to physicians. For further information, please see https://www.tga.gov.au/alert/medtronic-synchromed-ii-implantable-infusion-pump-0 .

Device

  • 모델명 / 제조번호(시리얼번호)
    Medtronic SynchroMed II PumpsModel Numbers: 8637-40 and 8637-20Manufactured before July 2011ARTG Number: 97770
  • Manufacturer

Manufacturer

  • 제조사 모회사 (2017)
  • 제조사 의견
    “If our surveillance systems identify a potential performance issue, our personnel promptly evaluate the problem, including, when appropriate, conducting root cause investigations and internal testing to assess whether the product continues to meet specifications and defined performance criteria,” Medtronic told ICIJ in a statement. “In some cases, based on this evaluation, Medtronic may determine that a recall is necessary.” The company said that it communicates with healthcare providers and/or patients and provide recommendations to address such issues. Medtronic noted that these communications can include letters, emails, calls, press releases, physician notifications and social media postings, as well as informing the FDA and other regulators of the actions.
  • Source
    DHTGA