Merge iConnect Enterprise Archive (iCEA) with software versions: 5.30.0 and greater up to and including 9.4.3(a picture archiving and communications system) 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Emergo Asia Pacific Pty Ltd T/a Emergo Australia 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2016-RN-00873-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2016-06-30
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    This recall has been initiated because the impacted versions of icea allow images to be stored to the icea without pixel data. - image was retrieved by radsuite from icea, a qc action was performed, & sent back to icea. - image was sent to icea with pixel data but the pixel data was not stored with the image to icea. the investigation found that there is a networking scenario where the network dropped the connection after sending the header information but before the pixel data arrived. icea stored the image without the pixel data.- image was sent to icea without pixel data & the image was stored to icea. the pixel data may have been lost by the device that sent the data to icea.A side effect of images without pixel data is performance degradation caused by the system continuously attempting to route the images to the peer system that fail to store. there could be delay in treatment in determining the progression of disease.
  • 조치
    Emergo is advising users that there is no work around for the issue, however the error is obvious to users using PACS viewer, as it is indicated that there is a missing image (slice) from the study. Users not using PACS viewer may not be notified of the error. Emergo will be implementing a software upgrade as a permanent correction. This action has been closed-out on 22/02/2017.

Device

  • 모델명 / 제조번호(시리얼번호)
    Merge iConnect Enterprise Archive (iCEA) with software versions: 5.30.0 and greater up to and including 9.4.3(a picture archiving and communications system)ARTG Number: 206683
  • Manufacturer

Manufacturer