Merge iConnect Enterprise Archive (iCEA) with software versions 8.30.7 through 9.4.3, with tier 2 storage enabled (a picture archiving and communications system) 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Emergo Asia Pacific Pty Ltd T/a Emergo Australia 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2016-RN-00864-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2016-06-30
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    This recall has been initiated due to an issue in icea which could result in unrecoverable loss of data when using tiered storage on icea. the issue can arise if a study has purged from tier 1 (i.E. via high water mark) and only resides in tier 2 and then the study is requested to be updated.In a tiered storage environment older data (less frequently accessed data) is staged to a second storage tier for long term storage. this staging process frees up space on the first storage tier for newer data. in rare circumstances, where the data that exists only on the second tier needs to be accessed, a software issue could lead to data being purged from both the first and second tier resulting in data loss. this issue may result in an archived study not being available for future viewing (i.E. as a prior study for comparison purposes, etc.). if prior studies are not available for viewing, this may lead to delay in treatment in determining the progression of the disease.
  • 조치
    Emergo is advising users to create a new object and remove the reference to the original version in order to prevent the issue from occurring. Emergo will be implementing a software update as a permanent correction. This action has been closed-out on 22/02/2017.

Device

  • 모델명 / 제조번호(시리얼번호)
    Merge iConnect Enterprise Archive (iCEA) with software versions 8.30.7 through 9.4.3, with tier 2 storage enabled (a picture archiving and communications system)ARTG Number: 206683
  • Manufacturer

Manufacturer