Merge PACS with software versions 6.5.5, 6.5.6, 6.5.7, 6.5.8, 6.5.9, 6.6, 6.6.1, 6.6.1.1, 6.6.2 (picture archiving and communication system) Distributed between June 2008 and June 2015 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Emergo Asia Pacific Pty Ltd T/a Emergo Australia 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2016-RN-00863-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2016-06-29
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    This recall has been initiated due to merge pacs inability to display presentation states on prior comparison studies generated on 3rd party systems or that have been migrated from office pacs. these presentation states are known as foreign presentation states. pacs 6.5.5 introduced the feature to be able to display foreign presentation states. however this feature was only available on the primary study in any given viewer instance. it did not apply to prior comparison studies that were opened in the same viewer. users might assume that if the annotations are available in the primary study, they would also be shown if available on the prior comparison studies.Use of this product may cause a user to believe that there was nothing significant seen in the prior image which may result in improper treatment. there have been no injuries reported as a result of this issue.
  • 조치
    Emergo are advising users to open prior comparison studies in the secondary viewer to check if there were annotations and read the report, or visit notes pertaining to that study to see if there are any noted findings that are otherwise not apparent. Emergo will be implementing a software upgrade as a permanent correction. This action has been closed-out on 23/05/2017.

Device

  • 모델명 / 제조번호(시리얼번호)
    Merge PACS with software versions 6.5.5, 6.5.6, 6.5.7, 6.5.8, 6.5.9, 6.6, 6.6.1, 6.6.1.1, 6.6.2 (picture archiving and communication system)Distributed between June 2008 and June 2015ARTG Number: 191068
  • Manufacturer

Manufacturer