microTargeting Depth Stop Adapter; microTargeting STar Drive System (Manual); microTargeting STar Drive System (Motorised); microTargeting Drive System; microTargeting STar Drive System (Manual) 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Emergo Asia Pacific Pty Ltd T/a Emergo Australia 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2018-RN-00663-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2018-06-07
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    During a dbs lead implant procedure, the fhc depth stop adapter is placed onto the lead to set the desired distance to target. the depth stop adapter then mounts in the lead measurement fixture to ensure that the lead is not inserted beyond the targeted depth. the fhc depth stop adapter may cause damage to the lead and stylet when the depth stop screw is over-tightened onto the lead. this damage can manifest as low impedances or a short circuit between one or multiple electrode pairs in the lead, and has resulted in the need to remove and replace the implanted lead intraoperatively or in a follow-up procedure. recent reports of lead short circuit events have led to an improvement of the depth stop design and a recall the previous version of this product. to date, there have been no reports of patient death related to this problem.
  • 조치
    Emergo is advising FHC and Medtronic Field Representatives will review user's inventory and replace affected Depth Stop Adapters on hand with an improved design. In the interim, Emergo advise that users do not over-tighten the Depth Stop Adapter screw. If low impedances or a short circuit is found, physicians should replace the lead prior to use and report the issue to FHC, Emergo Australia, or their local field rep.

Device

  • 모델명 / 제조번호(시리얼번호)
    microTargeting Depth Stop AdapterProduct Model Number: 66-AC-DS(1.8)microTargeting STar Drive System (Manual)Product Numbers: ST-DS-MA and 70-ZD-MAmicroTargeting STar Drive System (Motorised)Product Numbers: ST-DS-ME and 70-ZD-MEmicroTargeting Drive SystemProduct Numbers: FC1006 and 9033G0601microTargeting STar Drive System (Manual)Product Number: FC8001ARTG Number: 177545(Emergo Asia Pacific Emergo Australia - Stereotactic surgery system, neurological)
  • Manufacturer