Mindray's V-Series Patient Monitors, software versions 2.4.0.18, 2.4.0.50, 2.4.0.60 and 2.4.1.18 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Ulco Engineering Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2014-RN-01173-1
  • 사례 위험등급
    Class I
  • 사례 시작날짜
    2014-10-30
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    Mindray has identified an issue that involves the v-series drug calculator function specific to weight based calculations. this issue may occur when the patient’s weight is entered using either the v-series patient information dialog or the panorama central station (if it's in use with the central station). if the weight is manually entered, the issue will not occur. depending on the version of software in use, the most recent weight entered into the v-series patient setup dialog or the panorama central station may not be received by the calculator feature; the calculator may retain the original weight entered. additionally, the weight that is displayed in the drug calculator dialog may not reflect what was used for the calculation.If an incorrect dosing calculation results and medication is administered, immediate and/or long term health consequences may result depending on drug, dose, and duration.There are no reports of injuries associated with these issues.
  • 조치
    Mindray is recommending that customers discontinue use of the Drug Calculator function until the software correction is performed. This action has been closed-out on 26/08/2016.

Device

  • 모델명 / 제조번호(시리얼번호)
    Mindray's V-Series Patient Monitors, software versions 2.4.0.18, 2.4.0.50, 2.4.0.60 and 2.4.1.18 Model Numbers: 1412 & 1414Serial Numbers: EGL19008969, EGL1B009996, EGL1B010000, EGL1B010014, EGL1B010016, EGL1B010341, EGL1B010352, EGL1C010807ARTG Numbers: 223384
  • Manufacturer

Manufacturer