Mobile Detector Holder Trolley (9890-010-88152) (used to support the transport and utilisation of a portable/mobile radiological imaging system) 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Philips Electronics Australia Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2017-RN-01217-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2017-09-27
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    Philips has recently performed evaluations of the mobile detector holder and has identified a potential issue that may affect the performance of the equipment under certain conditions. if the lock mechanism of the movable clamp of the mobile detector holder is not adjusted properly or not secured, the portable detector can drop down when the moveable clamp is at the bottom (below of the detector) and additional force is applied. this may happen when the detector is rotated clockwise from landscape to portrait position.The wireless portable detector (wpd) is fixed in the mobile detector holder via a lock mechanism. the 4 countersunk screws, that are used to attach the lock mechanism, can become loose. if this happens and the mechanism fails, the detector may fall down when the holder is rotated 90° clockwise from landscape.
  • 조치
    1. Users should read the notice and check the lock mechanism of the Mobile Detector Holder. The lock mechanism can be moved without pressing the actuator, the holder should not be rotated clockwise. a. Distribute the notice to all users of the device and complete the supplied response form and return it to. Philips via fax/email within three (3) business days. b. Philips will contact users to schedule a correction to the affected units.