Monoclonal Mouse Anti-Human Leukaemia, Hairy Cell Clone DBA.44. An in vitro diagnostic medical device. 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Agilent Technologies Australia Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2018-RN-00329-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2018-04-16
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    It has been identified that the primary labels of the affected vials of m0880 were mislabelled with an incorrect concentration. the concentration was written to be 187 mg/l. the correct concentration is 485 mg/l.To date, no customer complaints have been received as a result of this issue.
  • 조치
    Agilent Technologies is advising customers to discard any bottles of the affected lots of this product in line with the precautions outlined in the IFU. Agilent will provide replacement product to affected sites who hold unused or partially used vials. Agilent Technologies is advising users who may have run results at an altered dilution due to the larger protein concentration and without appropriate run controls to review results and re-run where required.

Device

  • 모델명 / 제조번호(시리얼번호)
    Monoclonal Mouse Anti-Human Leukaemia, Hairy Cell Clone DBA.44. An in vitro diagnostic medical device.Code Number: M0880 Lot Numbers: 20033445 and 20042272ARTG Number:183436(Agilent Technologies Australia - Immunohistology cell marker IVDs)
  • Manufacturer

Manufacturer