Monopolar HF Resection Electrodes 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Olympus Australia Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2017-RN-00498-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2017-04-20
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    The manufacturer, olympus winter & ibe has received an increased number of complaints regarding loop wires breaking at the distal end of the electrodes. investigations have confirmed that loop wires can break during the intended use of the electrodes. as a result, a fragment may fall inside the patient and will need to be retrieved. under certain circumstances, the retrieval could require additional surgical treatment. to date, no reports of adverse events or patient injuries have been received by olympus.
  • 조치
    Olympus is requesting users to inspect their inventory and quarantine the affected lots. Affected products are to be returned for replacement.

Device

  • 모델명 / 제조번호(시리얼번호)
    Monopolar HF Resection ElectrodesModel Number: A22201CLot Numbers: 16123P03L001, 16139P03L001, 16144P03L001, 16173P03L001, 16174P03L001, 16210P03L001, P16X0001, P16X0002, P16Y0001 and P16Y0003Model Number: WA22037CLot Number: P16Y0001ARTG Number: 218223