MOSAIQ 2.62 and higher with the IQ Scripts license enabled (Oncology information system used to manage workflows for treatment planning and delivery) 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Elekta Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2016-RN-00781-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2016-06-14
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    Clicking on the treat button in the treatment chart window will trigger the plan-of-the-day iq script to run. the plan-of-the-day window will open and the treatment plan can be selected by clicking on the radio button. it is possible that the selected plan-of-the-day could be accidentally changed by scrolling the mouse wheel up and down and inadvertently selecting another plan.Though the wrong plan-of-the-day selected for one fraction would likely not result in critical harm, if the wrong plan were selected multiple times it could result in serious injury.
  • 조치
    The issue is corrected in the general release of MOSAIQ 2.64. A software patch will be released for MOSAIQ version 2.62 in order to correct the defect. In the interim, the problem can be avoided by careful use of a mouse with a wheel (or using a mouse without a scroll wheel) and verifying that the correct plan is chosen before selecting OK on the Plan-of-the-Day form. This action has been closed-out on 02/02/2017.