MOSAIQ version 2.5 and above - Medical Oncology only 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Elekta Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2015-RN-01216-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2015-12-10
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    There is a possibility of incorrect drug dosage calculation due to the following software/data item issues:when there is no valid (unexpired) patient weight entered on the same date or prior to a serum creatinine value being populated, a stale creatinine clearance (crcl) value is being used in area under curve (auc) dose calculations. as a separate issue, if there is a valid age limit added to an elekta metric data item (e.G. weight (kg)) and the same valid age limit has not been added to the elekta imperial data item which corresponds (e.G. weight (lb)), the calculated result which has no valid age will still be considered valid and could be used in drug dose calculations.There is a remote probability that critical harm could occur. no patient injuries have been reported so far.
  • 조치
    The sponsor is advising users to implement the following workaround prior to the installation of a software update: always enter a current weight for the patient who is receiving Carboplatin on the same day or prior to the serum creatinine value being populated; Add the valid age in all Elekta data items.

Device

  • 모델명 / 제조번호(시리얼번호)
    MOSAIQ version 2.5 and above - Medical Oncology onlyARTG Number: 225164
  • Manufacturer

Manufacturer