MultiDiagnost Eleva and MultDiagnost Eleva with Flat Detector (diagnostic fluoroscopic x-ray system) 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Philips Electronics Australia Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2016-RN-00943-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2016-07-15
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    The two bolts that attach the clamping plate to the lateral shaft may break. this may cause the c-arm to float freely. depending on the orientation of the free-floating c-arm and any other forces applied to it, the c-arm could move laterally forward, slide out of its bearings and then possibly sag. the sagging c-arm can strike a patient or operator. however, the probability of this occurring is rare. there have been no reports of injury associated with this event.
  • 조치
    Philips is advising users to be aware of the issue. If the C-arm moves in an unexpected manner, users should contact Philips immediately. Philips will be implementing a repair of the C-arm as a permanent correction. This action has been closed-out on 28/06/2017.

Device

  • 모델명 / 제조번호(시리얼번호)
    MultiDiagnost Eleva and MultDiagnost Eleva with Flat Detector (diagnostic fluoroscopic x-ray system)MultiDiagnost ElevaProduct Code: 708032MultiDiagnost Eleva with Flat DetectorProduct Code: 708034MultiDiagnost ElevaProduct Code: 708036MultiDiagnost Eleva with Flat DetectorProduct Code: 708037 MultiDiagnost Eleva with Flat DetectorProduct Code: 708038ARTG Number: 98560
  • Manufacturer

Manufacturer

  • 제조사 모회사 (2017)
  • Source
    DHTGA