MX 16-slice CT x-ray systems with software version v1.1.4.21426. 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Philips Electronics Australia Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2016-RN-00616-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2016-05-12
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    The following issues have been found in mx 16-slice systems.1) during the filming operation, the clipboard used for copying and pasting images is not cleared between patients. if the operator fails to copy the current patient’s images before pasting, a previous patient’s images may be present in the clipboard and be copied onto the film of the current patient. 2) during the bolus tracking scan, if the auto voice in tracker scan is enabled, the tracker scan will be aborted unexpectedly and the diagnostic scan after the tracker scan needs to be manually started.3) if the scan protocol with sas (spiral auto start) function is selected to plan the scan, the sas option may not be displayed on the contrast tab of the scan protocol parameter area or the sas option is displayed but not enabled as pre-configured in the protocol. this issue only occurs on the first helical scan after system startup that applies sas function.There have been no reports of injuries.
  • 조치
    Philips is providing users with work around instructions to follow as an interim measure. A software upgrade will be implemented as a permanent correction. This action has been closed-out on 02/03/2017.

Device

  • 모델명 / 제조번호(시리얼번호)
    MX 16-slice CT x-ray systems with software version v1.1.4.21426.ARTG Number: 158775
  • Manufacturer

Manufacturer

  • 제조사 모회사 (2017)
  • Source
    DHTGA