N Latex Carbohydrate-deficient Transferrin (CDT). An in vitro diagnostic medical device (IVD). 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Siemens Healthcare Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2017-RN-00452-1
  • 사례 위험등급
    Class III
  • 사례 시작날짜
    2017-04-07
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    Siemens healthcare diagnostics has observed lower values for carbohydrate-deficient transferrin (cdt) measurement when using the affected n latex cdt lots in comparison to former lots. cdt results in absolute concentrations are influenced by the patient’s total transferrin levels and therefore results are reported as ratio of cdt to total transferrin, called %cdt. as the transferrin determination with n antiserum to human transferrin is not affected, the lower cdt result leads to calculated %cdt values that show a negative bias of approximately 10 to 15% compared to former lots. this could result in a shift of weak positive patient samples into the reference range of 1.19 – 2.47 %cdt which was derived from a study population of healthy adults. it is not expected that samples from patients with chronic heavy alcohol consumption will be misclassified.
  • 조치
    Siemens is advising users to discontinue use of and discard the affected lots. Replacement stock will be provided by Siemens. It is recommended that users review the customer letter with the laboratory Medical Director.

Device

  • 모델명 / 제조번호(시리얼번호)
    N Latex Carbohydrate-deficient Transferrin (CDT). An in vitro diagnostic medical device (IVD).Catalogue Number: OPCS035Siemens Material Number: 10445996Lot Numbers: 45834 & 45987 (Expiry 3 Jun 2017); 46116, 46244, 46388 & 46568 (Expiry 10 Oct 2017)ARTG Number: 178153
  • Manufacturer

Manufacturer