N Latex ß2–Microglobulin 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Siemens Healthcare Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2015-RN-01054-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2015-11-02
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    Siemens healthcare diagnostics has received complaints from customers not being able to establish a calibration curve or to validate an established calibration curve with the respective controls when using one of the affected lots of n latex ß2-microglobulin. the issue might lead to the inability to calibrate the assay or to obtain invalid test runs. however, patient results are not affected. there is the potential for the user to be exposed to assay components contaminated with pseudomonas libanensis, pseudomonas rhodesiae and pseudomonas extremorientalis. exposure may lead to an infection and its consequencesa review of previously generated results due to this issue is at the discretion of the laboratory.Siemens advises that the delay of results for these assays does not represent a hazard, and as a result the severity level is none or negligible.
  • 조치
    Customers are asked to discontinue use of and discard the affected lots listed in the Customer Letter. A replacement or credit will be issued as per the normal procedure.

Device

  • 모델명 / 제조번호(시리얼번호)
    N Latex ß2–Microglobulin Catalogue Numbers: OQWU155 and OQWU15 Lot Numbers: 44862, 44902, 44968, 45192, 45229, 45252, 45291, 45329, 45342, 45366, 45470, 45525 and 45543. ARTG Number: 178153