Natus NicoletOne 32Ch Ambulatory EEG Record System 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Life Healthcare Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2018-RN-00141-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2018-02-27
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    The supplier, natus has discovered a scenario that incorrectly allows customers using nicvue to import an ambulatory exam into a patient for whom the exam was not started. in the event that after starting an ambulatory exam, the recorder application does not shutdown correctly, an empty exam file is left in the patient exam list in nicvue. with this empty exam file present, no internal check is done to ensure the patient id on the ambulatory’s sd card matches the selected patient in nicvue, resulting in the exam on the sd card being imported into the patient record that is currently selected in nicvue. the register.Exe application will bypass the patient id check on the card of the selected patient and allow the exam to be registered into any selected patient.
  • 조치
    LifeHealthcare is advising users to ensure the recorder is only shutdown via the Recorder UI, either from the File menu or the X button. If the issue occurs, the empty exam can be deleted from NicVue prior to importing the ambulatory exam. LifeHealthcare will be conducting training for relevant staff regarding this issue, and providing users with an updated copy of the Instructions for Use.

Device

  • 모델명 / 제조번호(시리얼번호)
    Natus NicoletOne 32Ch Ambulatory EEG Record SystemSerial Number: MBR160952ARTG Number: 294968
  • Manufacturer

Manufacturer