Neuromuscular Transmission Module used with Carescape and Datex-Ohmeda S/5 Anaesthesia monitors 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 GE Healthcare Australia Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2014-RN-00179-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2014-02-19
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    When the e-nmt-01 module is used in conjunction with the electrosensor, the neuromuscular transmission (nmt) values may indicate a deeper level of muscle relaxation than the actual level of muscle relaxation. in the clinical situation visual movements of the hand are seen after tof (train of four) stimulation, but the patient shows no counts, or counts are not corresponding to the actual amount of movements. this issue may lead to an inadequate dose of muscle relaxants. this issue may occur when the e-nmt-01 module is plugged into the carescape or the datex-ohmeda s/5 anesthesia monitor. module e-nmt-00 is unaffected by this issue.
  • 조치
    GE Healthcare is advising users to do not use the E-NMT-01 module with ElectroSensor. The E-NMT-01 module can continued to be used with the MechanoSensor and Paediatric MechanoSensor. A software update will be provided to all affected users as a permanent correction. This action has been closed-out on 10/02/2016.

Device

  • 모델명 / 제조번호(시리얼번호)
    Neuromuscular Transmission Module used with Carescape and Datex-Ohmeda S/5 Anaesthesia monitors Module Version: E-NMT-01ARTG Number: 140216