Nitinol RF Reusable Electrodes 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Life Healthcare Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2017-RN-00846-1
  • 사례 위험등급
    Class I
  • 사례 시작날짜
    2017-07-11
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    The manufacturer, cosman medical, has received complaints regarding nitinol rf (model tcn) reusable electrodes. after multiple reprocessing cycles, the epoxy resin which holds the tcn electrode in the hub can exhibit signs of damage. this damage may result in the inability to fully remove blood and/or tissue residuals prior to cleaning and re-sterilising the device. the failure to completely remove any blood and/or tissue from the tcn electrodes prior to re-sterilisation creates a potential risk of infection and spread of blood borne pathogens to patients.Cosman medical is implementing an action for nitinol rf (tcn) reusable electrodes to convert them to single patient use electrodes (re-usable for one patient only). after single patient use, the used tcn electrode should be disposed of in accordance with the individual institution’s infectious material/biohazard waste control procedures. to date, cosman have not received any reports of patient injury related to this issue.
  • 조치
    Action to be taken by users: Cosman Medical is implementing an action for Nitinol RF (TCN) reusable electrodes to convert them to single patient use electrodes. After single patient use, the used TCN Electrode should be disposed of in accordance with the individual institution’s infectious material/biohazard waste control procedures. Unopened TCN Electrodes need to be identified as single patient use only (re-usable for one patient only). These devices are supplied non-sterile which require cleaning and sterilisation prior to use. These TCN electrodes may be used once the cleaning and sterilisation steps are completed. Isolate any used affected items. A LifeHealthcare Representative will arrange collection and replacement. Return the Reply Form provided by LifeHealthcare so they can validate that all customers have been advised of this action.

Device

  • 모델명 / 제조번호(시리얼번호)
    Nitinol RF Reusable ElectrodesProduct Codes: TCN-5, TCN-5-3M, TCN-10, TCN-10-3M, TCN-15, TCN-15-3M, TCN-20, TCN-20-3M, TCNK-5, TCNK-5-C, TCNK-5-R, TCNK-10, TCNK-10-C, TCNK-10-R, TCNK-15, TCNK-15-C, TCNK-15-R, TCNK-20, TCNK-20-C, TCNK-20-RAll Lot NumbersARTG Number: 158235
  • 의료기기 분류등급
  • Manufacturer

Manufacturer