Nitinol Staples Elastic Staples - Large and Small Symmetrical 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 LMT Surgical Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2018-RN-00354-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2018-04-26
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    During a retrospective review of sterilisation records, trimed has determined that evidence of quarterly dose audits/verification required to comply with the relevant regulations could not be found. sterilization records indicate that all lots of staples were irradiated to at least the defined dose and although trimed has not received any reports of adverse events or injuries, there is a risk that this issue could result in patient infection if not properly sterilised.
  • 조치
    As a permanent corrective action, TriMed will recall all of the Nitinol Staples that were manufactured between March 12, 2015 to the present date.