NovoPen Echo 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Novo Nordisk Pharmaceuticals Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2017-RN-00832-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2017-07-10
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    Novo nordisk a/s has detected that the insulin cartridge holder used in a small number of novopen echo batches may crack or break after exposure to certain chemicals, in particular, certain cleaning agents. using a novopen echo insulin cartridge holder that is cracked or broken could result in the device delivering a smaller dose of insulin than expected, which can lead to serious health consequences including hyperglycaemia (high blood sugar).There are no reported problems with the insulin being administered by these devices and that cracking and breaking should not occur if the device is cleaned in accordance with the instructions in the user guide. the risk of patients experiencing problems as a result of this issue is considered extremely low.
  • 조치
    Patients: Where affected pens have been issued to a patient, Novo Nordisk urges patients to replace the cartridge holder, as some could be damaged (as per User Guide). Register your contact details (name, address, phone number, email and number of affected cartridge holders) either at www.novonordisk.com.au or www.novonordisk.com/novopenecho5.html in order to receive a replacement cartridge holder for your NovoPen Echo (after 7 working days, approximately). Health Care professionals: Please immediately quarantine all affected batches within your possession to prevent further distribution. Complete and return the Acknowledgment form even if no affected stock is held. Whenever the opportunity presents, please reiterate to patients that the cleaning instructions of the NovoPen Echo User Guide (page 26) describe that a cloth dampened ONLY with water (and certainly not with bleaching agents such as chlorine, iodine or alcohol) is to be used to clean the insulin pen.

Device

  • 모델명 / 제조번호(시리얼번호)
    NovoPen EchoBatch Numbers (as shown on NovePen Echo Carton): EVG3310-6, EVG4252-2, EVG4253-2, FVG8218-4, FVG8412-3 and FVG8415-1Batch Numbers (as shown on NovePen Echo Pen): EVG3310, EVG4252, EVG4253, FVG8218, FVG8412 and FVG8415ARTG Number: 189785