O2Vent T(Mandible-repositioning anti-snoring orthosis) 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Oventus Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2016-RN-01554-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2016-12-05
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    The manufacturer has determined that a correction to the instructions for use (ifu) in relation to the adjustment range for titration is necessary. the adjustment of the screw and hook allows for appropriate titration by mandibular advancement for optimised treatment. the therapeutic treatment is unaffected by the revised adjustment range for titration. the revised adjustment range is as follows:one full turn using the adjustment key is 360 degrees. this will adjust the pitch of the screw by 0.5mm resulting in a 0.5mm titration. the screw should be adjusted a maximum of 1mm per night (i.E. 2 full turns of the adjustment key).This update only affects some patients who are currently being reviewed for optimised treatment.
  • 조치
    Clinicians are advised that if their patient’s treatment has been optimised with the current positioning of their device, this correction does not affect them. They are also advised to provide a copy of the updated Instructions of Use to their patients. This action has been closed-out on 26/05/2017.

Device

  • 모델명 / 제조번호(시리얼번호)
    O2Vent T(Mandible-repositioning anti-snoring orthosis)Catalogue Number: O2VTDLARTG Number: 275667