OER-AW Endoscope Reprocessor 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Olympus Australia Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2017-RN-00130-1
  • 사례 위험등급
    Class III
  • 사례 시작날짜
    2017-01-30
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    There is an update of the operation manual for the oer-aw endoscope reprocessor to instruct users of non-detachable distal cover type duodenoscopes, such as the tjf-q180v duodenoscope, to follow the pre-cleaning and manual cleaning as stated in the duodenoscope’s reprocessing manual prior to setting them into the oer-aw.The following warning has been added to the list of compatible endoscopes/connecting tubes and chapter 4 “basic endoscope reprocessing operations” of the oer-aw operation manual - "when reprocessing the following non-detachable distal cover type duodenoscopes using oer-aw, conduct pre-cleaning and manual cleaning as detailed in each duodenoscope’s reprocessing manual".
  • 조치
    Olympus is requesting users to inspect their inventory and replace any existing OER-AW Operation Manuals and List of Compatible Endoscopes/Connecting Tubes with the latest version supplied.They are further advised to implement the instructions enclosed in the updated OER-AW Operation Manual. This action has been closed-out on 06/07/2017.

Device

  • 모델명 / 제조번호(시리얼번호)
    OER-AW Endoscope ReprocessorImpacted models of duodenoscopes: JF-20, JF-140F, TJF-M20, TJF-30, TJF-145, TJF-140F, TJF-150, TJF-160F, TJF-160VF and TJF-Q180VARTG Number: 185051
  • Manufacturer

Manufacturer