Offset Cup Reamer Handle(used to hold both the Registration Array during Registration and the Acetabular Reamer Basket during acetabular reaming in total hip procedures) 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Stryker Australia Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2016-RN-01468-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2016-11-09
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    Stryker has received reports of two potential issues related to the offset cup reamer handle. the weld on the u-joint of the handle may become damaged over time, potentially causing the pin or pieces of the pin to fall into the wound. if that occurs, and the surgery proceeds with reaming, the pin or pieces of the pin may fall into the wound, which may be removed if visible or may be left in the wound. the handle also may not fully engage with the registration array and/or reamer basket. if that occurs, the surgery may proceed using a backup offset reamer handle, a straight reamer handle or manual instrumentation.
  • 조치
    Customers are advised to inspect their inventory for affected product and quarantine it immediately. A Stryker representative will coordinate the return of any affected products. This action has been closed-out on 28/06/2017.

Device

  • 모델명 / 제조번호(시리얼번호)
    Offset Cup Reamer Handle(used to hold both the Registration Array during Registration and the Acetabular Reamer Basket during acetabular reaming in total hip procedures)Item Number: 207080All Lot Numbers affectedContained within trays: 151550, 151550-TARTG Number: 223107
  • Manufacturer

Manufacturer

  • 제조사 모회사 (2017)
  • Source