Omega+ Lag Screw Ø13X110MM ST 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Stryker Australia Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2017-RN-00526-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2017-04-28
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    Stryker gmbh has received a customer complaint that the lag screw would not completely cannulate as the guide wire would not pass through the cannula of the screw. no harm was reported. it was found on investigation that the obstruction is related to a metal burr in the cannula of the screw which was not removed during manufacturing causing obstruction. the obstruction in the cannula of the screw is easily recognisable for the user. in most cases the surgeon would detect significant resistance when passing the lag screw over the guide wire before it touches the skin and would not continue using the screw. a different screw would be available and used, resulting in a minor delay in surgery of <15 minutes. if the surgeon does not notice or ignores the resistance, then surgical complications related to the guide wire being stuck in the cannulated screw and/or the guide wire being pushed out of the femoral head could theoretically occur.
  • 조치
    Stryker is advising customers to immediately inspect inventory and quarantine affected product. Stryker will coordinate the return and replacement of the product.

Device

  • 모델명 / 제조번호(시리얼번호)
    Omega+ Lag Screw Ø13X110MM STItem Number: 33625110Lot Number: V06279Ø13X105MM STItem Number: 33625105Lot Number: V06269ARTG Number: 119660
  • Manufacturer

Manufacturer

  • 제조사 모회사 (2017)
  • Source
    DHTGA