Omrix Pressure Regulator (Surgical air pipeline regulator) 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Johnson & Johnson Medical Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2014-RN-00937-1
  • 사례 위험등급
    Class I
  • 사례 시작날짜
    2014-08-26
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    Omrix biopharmaceuticals has identified that there may be pressure regulators in the field with an incorrect connector type. a nitrous oxide connector has been supplied instead of a carbon dioxide connector (sis-co2 connector). the connector has a different colour (grey and not green-grey, as should be according to the australian standard- as 2896- for medical gas systems ) and in addition, instead of carbon dioxide, it was engraved with “nitrous oxide’’. johnson & johnson medical pty limited (jjm) is recalling omrix pressure regulators. the nitrous oxide inlet connector is not compatible with carbon dioxide source connector and can therefore not connect to omrix pressure regulator.
  • 조치
    Johnson & Johnson Medical Pty Limited is requesting the customers to quarantine the affected pressure regulators. Arrangements will be made for a courier to pick up the affected product and a credit note issued. This action has been closed-out on 22/03/2016.

Device

  • 모델명 / 제조번호(시리얼번호)
    Omrix Pressure Regulator (Surgical air pipeline regulator)Product Codes: EVRC68 and EVR54CARTG Number: 203072
  • Manufacturer

Manufacturer