Omrix Pressure Regulators (used to spray Evicel Solutions for Fibrin Sealent during open and laproscopic procedures to prevent bleeding) 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Johnson & Johnson Medical Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2016-RN-00320-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2016-03-22
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    The european medicines agency (ema) has clarified their recommendation for 20 psi (as opposed to 20-25 psi) pressure to be used when spraying evicel with the 45cm tip during laparoscopic surgery. as a result, the labelling of omrix pressure regulators in australia need to be updated and the sticker label on each pressure regulator will be replaced to reflect this change.Also, in an effort to enhance consistency between the labelling of the omrix pressure regulator and the omrix evicel solutions for sealant, the ifu for the pressure regulator has been updated to remove the following statement;"spray application of evicel should not be used in endoscopic procedures where the minimum recommended distance of 4 cm cannot be maintained. for laparoscopic procedures using accessory tips, see the recommendations in table 2".
  • 조치
    Johnson & Johnson Medical (JJM) is providing users with the updated IFU and replacing stickers on the Omrix Pressure Regulators to reflect the new recommendation of 20psi when spraying Evicel with a 45cm tip during laparoscopic surgery. This action has been closed-out on 30/01/2017.

Device

  • 모델명 / 제조번호(시리얼번호)
    Omrix Pressure Regulators (used to spray Evicel Solutions for Fibrin Sealent during open and laproscopic procedures to prevent bleeding)Product Number: EVRC68 AND EVR54CARTG Number: 203072
  • Manufacturer

Manufacturer