Oncentra External Beam (formerly known as Oncentra MasterPlan during release of v3.x)(Radiation therapy treatment planning software) 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Nucletron Pty Limited 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2016-RN-01655-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2016-12-22
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    If a physical wedge, for which the total field size limit perpendicular to the wedge direction is more than 10 cm larger than the field size limit in wedge direction, is inserted in in-plane direction (as seen with collimator angle 0 degrees), then with very large fields perpendicular to wedge direction a part of the scattered fluence can become unmodulated by the wedge. this will result in profile “horns” close to the edge of the field. the deviation is most prominent with the varian 60-degree wedge. the calculated overdose in the affected areas, towards the edges of the field, is typically 6% for clinically used field sizes locally. smaller deviations inside the field can also be observed with the varian 45-degree wedge. with all other inplane wedges, including the elekta motorised wedge, there is only an insignificant deformation of the cross profile tail, which is outside the treatment field.
  • 조치
    Elekta will be correcting the issue in the next upgrade of the Oncentra External Beam. In the interim, users are advised to not exceed jaw positions of +/- 12cm with the Varian 60-degree wedge and +/- 14.5 cm with the Varian 45-degree wedge in the direction perpendicular to the wedge slope, if the wedge is inserted in in-plane direction (collimator angle 0 degrees). In addition, Elekta is strongly advising users to perform proper Quality Assurance for all treatment plans before delivery of the first fraction to the patient.

Device

  • 모델명 / 제조번호(시리얼번호)
    Oncentra External Beam (formerly known as Oncentra MasterPlan during release of v3.x)(Radiation therapy treatment planning software)All versionsARTG Number: 142110
  • Manufacturer

Manufacturer