Optima CT660, Brivo CT385, Discovery PET/CT 610 and Discovery PET/CT 710 Systems (when used with certain PACS systems) 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 GE Healthcare Australia Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2013-RN-00326-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2013-04-16
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    There is an issue related with the direct multi-planar reformat (dmpr) function and screen saved images on the scanner. the dmpr and screen save functionality may possibly generate the same series uid. it has been seen that some picture archiving and communication system (pacs) may combine series under a patient using only the series uid, although all images are correctly annotated. this issue can cause additional images from a different patient and images from two or more patients to appear in one exam. please note that this issue has only been seen on pacs systems manufactured by intelerad (intelepacs 4-3-1-p350 and intelebrowser -4-3-1-p350) at this time.
  • 조치
    GE Healthcare is recommending that customers discontinue using the DMPR functionality, if they are using the above PACS version or if they see similar issues from their current PACS and utilise an alternative capability, such as Batch-mode Reformat on the GE AW, AW server or scanner operator console or 3D reformat save on their 3D PACS viewer. GE Healthcare is planning to implement the permanent correction to affected systems when it is available.

Device

  • 모델명 / 제조번호(시리얼번호)
    Optima CT660, Brivo CT385, Discovery PET/CT 610 and Discovery PET/CT 710 Systems (when used with certain PACS systems)ARTG Number: 156650