Optima XR 220amx, Optima XR200amx with Digital Upgrade Mobile X-ray Systems 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 GE Healthcare Australia Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2013-RN-01321-1
  • 사례 위험등급
    Class I
  • 사례 시작날짜
    2013-12-13
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    Intermittently when a large number of patient exams are retrieved from the his/ris system, the exams when displayed on the mobile x-ray system are not properly displayed. if the user continues to attempt to select one of the exams when this issue has occurred, the system will update the screen and in doing so, potentially select a different exam/patient. if the user does not notice this change and continues to perform the exam, the images will be placed in an incorrect patient file.
  • 조치
    GE Healthcare is advising users to pay close attention to the patient exam/entry selected from the worklist prior to selecting the Start Exam button to ensure that the system did not update and select an incorrect exam. GE Healthcare is providing a software upgrade to permanently correct the issue.

Device

  • 모델명 / 제조번호(시리얼번호)
    Optima XR 220amx, Optima XR200amx with Digital Upgrade Mobile X-ray SystemsModel Numbers: 5555000-5, 5555000-6 and 5421698ARTG Number: 201698
  • 의료기기 분류등급
  • Manufacturer

Manufacturer